FT光谱仪测量调制光和脉冲光的应用实例/APPLICATION NOTE: MODULATED & PULSED LIGHT S OURCES

2022年12月15日 16:28:49      来源:北京锦坤科技有限公司 >> 进入该公司展台      阅读量:33

分享:

FT光谱仪测量调制光和脉冲光的应用实例/APPLICATION NOTE: MODULATED & PULSED LIGHT S OURCES

FT光谱仪测量调制光和脉冲光的应用实例/APPLICATION NOTE: MODULATED & PULSED LIGHT S OURCES

This application note describes the use of ARCspectro FT -NIR and FT-MIR for measuring modulated (or
pulsed) light sources
. It is shown that pulsed light sources with frequencies ranging from 10kHz to over
80Mhz can be measured with our systems.

I NTRODUCTION
The ARCspectro FT series are scanning Fourier-transform spectrometers. These instruments use a white -light interferometer to record an interferogram by scanning the position of one of the interferometer
mirrors. Nevertheless, time -varying light sources can be measured with the Arcspectro FT series , provided
that the frequency of the source lies outside the interfe rometer modulation frequenc y range. Typically, the
interfe rometer modulation frequencies are in the order of a few kHz. Furthermore, a digital filter is
implemented in the processing, which fil ters out any frequency component lying outside the 100 Hz - 10kHz
range (these are indicative values that might differ from one spectrometer type to another). Modulated or
pulsed light sources can be measured provided that their f requency (or repetition rate) lies above of the
higher interferometer modulation frequency (typically 10kHz) . Below are three measurement examples .

E XAMPLE 1 – P ULSED LED 1550NM
As a first example, a 2.5mW LED emitting broadly around 1. 55𝜇𝑚 was measured using a n ARCspectro FT -NIR equipped with a 0.9 -2.6 microns InGaAs detector. T he LED output was modulated at different
frequencies using a function generator .

Figure 1 shows 10 subsequent measured spectra of the LED modulated at 15 kHz . T his frequency is above
the interfe rometer modulation frequencies and internal digital fileter. T herefore, t he spectrum of th e LED is
measured normally.



EXAMPLE 2 - 80MH Z B ROAD- BAND SUPERCONTINUUM SOURCE
A pulsed supercontinuum source similar to, but not identical to, a NKT Photonics SuperK Extreme EXW -12
system was measured on a Arcoptix FT -MIR 2.0-6.0.

This supercontinuum source delivers light in ca 500 ps pulses with 80 MHz covering a spectrum of 450-2400
nm with an average power of about 5W. The spectrum was mea sured on the Arcoptix FT -MIR 2-6.0 with a
diffusion lens before the FT -MIR to reduce the intensity of the 1mm collimated beam from the
supercontin uum source. T he laser power was varies from 20 to 99%. Results are shown on Figure 3 .



上一篇:SpectroSolar - Example Measurements
下一篇:Comparison of the Arcoptix’s ARCSpectro HT with a standard grating spectrometer

版权与免责声明:
1.凡本网注明"来源:全球资源网"的所有作品,版权均属于全球资源网,转载请必须注明全球资源网。违反者本网将追究相关法律责任。
2.企业发布的公司新闻、技术文章、资料下载等内容,如涉及侵权、违规遭投诉的,一律由发布企业自行承担责任,本网有权删除内容并追溯责任。
3.本网转载并注明自其它来源的作品,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容的真实性,不承担此类作品侵权行为的直接责任及连带责任。其他媒体、网站或个人从本网转载时,必须保留本网注明的作品来源,并自负版权等法律责任。 4.如涉及作品内容、版权等问题,请在作品发表之日起一周内与本网联系。